Table 5. Comparison of the Ct value of E. sakazakii (Cronobacter spp.) and non-E. sakazakii (non-Cronobacter spp.) using by various real-time PCR machine
Type of real-time PCR machine | List of stains tested in this study (Ct value) | Volume of format |
Escherichia vulneris | Enterobacter cloacae | E. sakazakii (Cronobacter spp.) | Salmonella Enteritidis | Water | Positive control | Negative control |
ES626 | N6 | FSM282 | N6 |
ABI Prism 7000 SDS | Undet1) | Undet | 24.88 | 25.15 | 20.90 | 24.05 | Undet | Undet | 22.49 | Undet | 50 µL |
Rotor-gene | Undet | Undet | 17.03 | 23.59 | 15.90 | 17.25 | Undet | Undet | 15.76 | Undet | 25 µL |
Smart cycler | Undet | Undet | 18.33 | 23.23 | 17.22 | 19.55 | Undet | Undet | 17.59 | Undet | 25 µL |
Light cycler | Undet | Undet | Undet | Undet | Undet | Undet | Undet | Undet | Undet | Undet | 20 µL |
undetection.